Air-Cooled Temperature Test Chamber 335 Litre Climatic Test Chamber
This test chamber is suitable for high temperature,low temperature and temperature cycling test of the whole units,parts and materials.
Test Space | 335L | ||
Internal size | 580*765*750mm (W*D*H) | ||
External size | W870mm*D1860mm*H1800mm(W*D*H) | ||
Power supply | 230V±10%,50Hz±1,60A, 3/N/PE, earth resistance < 4Ω | ||
Max. Power | 9.5KW | ||
Max. Currency | 21A | ||
Cooling Method | Air-Cooled(Cooling system)/(Optional water cooling ) | ||
Installation Site |
|
Temp Test
| Temp.Range:-70℃~+180℃ | Note:
1.Test with no load, meeting GB/T5170 2008, if no other special notice.
2.Ambient Temp 25℃, good ventilation condition: function well if 35℃≥RT>25℃;Cool down rate and thermal load capacity will be affected by ambient low temperature.
| ||||||||
Temp Fluctuation:±2℃ | ||||||||||
Temperature Uniformity:±0.5℃~±2℃ | ||||||||||
Temp Deviation :±0.1℃~±0.5℃ | ||||||||||
Heat Up Time:<45min(Room Temp.<+180℃) | ||||||||||
Cooling Time:<50min(Room Temp→-70℃) | ||||||||||
Test standard | 1. GB/T 2423.1 Low Temp. Test | 2.GJB 150.3 High Temp. Test | ||||||||
3.GB/T 2423.2 High Temp. Test | 4.GJB 150.4 Low Temp. Test | |||||||||
|
Structure Mode | Unitary.Test room is on the top, refrigerator is in the bottom; Power distribution control in the back; Main control panel in the front door, and the extended one is on the right-back side. Door opens from right to left. Air circulation channel and air conditioning system are at the back of test space:(centrifugal fan blade,air flow guiding device,heater,humidifier,Refrigeration heat exchanger,Temp and humidity sensing device. |
Composition |
|
| |
| |
Load-bearing
|
|
| |
| |
Chamber door |
|
| |
| |
Observation window | Size:280×400mm(W×H), coated heating anti frost hollow glass |
Cable ports | 1 port with Φ100 mm on left and right sides, with screw caps and Soft silicone plugs. |
Headlamp | In the observation window |
Shelves | 2 shelves inside for samples; height adjustable |
Drain Holes | Condensate water drain holes for condensate water outside |
Spill Port | Located at the back of case for condensate water discharging |
Mobile Castors | Strong adjustable wheels |
Separate Design Of Water/Electricity | Humidity system and electric control system are separated for safety(patent no.:ZL201220245236.8) |