Brand Name :Envsin
Model Number :EYC605S2
Certification :CE and ISO
Place of Origin :China
MOQ :1
Price :To Be Negotiated
Payment Terms :L/C, T/T, Western Union
Supply Ability :25 per month
Delivery Time :30-35 business days after received the payment
Packaging Details :Wooden packing
Product name :Thermal Shock Chamber
Keywords :Thermal Shock Test Chamber
Test Space :2.6L
Specimen box Size :120*120*150mm
Liquid Tank Size :250*350*400mm
External size :1000mm*17500mm*1900mm
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

4.5L Liquid to Liquid Thermal Shock Chamber Fluorocarbon oil Thermal Shock Chamber

The thermal shock test chamber is suitable for examining the ability of the whole device, components and parts to withstand sharp temperature changes. The temperature impact test can understand the impact of the test sample once or continuously for many times due to temperature changes. The main parameters that affect the temperature change test are the high temperature and low temperature values in the range of temperature change, the sample holding time at high temperature and low temperature, and the number of test cycles.
Product Feature
Test Space 2.6L
Specimen box size

120*120*150mm

Liquid Tank Size

250*350*400mm

External size

1000mm*17500mm*1900mm

Thermal Shock Test Chamber for Rapid Temperature Shifts

Liquid test medium

Fluorocarbon oil

Power supply

AC380V±10%,50Hz±1

Maximum system power

15KW

Maximum system current

27A

Cooling Method Water-cooled (refrigerating unit)

Main technical parameters

Liquid temperature shock test parameters

Low temperature liquid impact test range

High temperature liquid impact test range

-65℃~0℃

+70℃~+150℃

Temperature fluctuation

±0.3℃~±1.0℃

Temperature uniformity

±0.5℃~±2.0℃

The temperature deviation

<±2℃

High temperature liquid tank liquid preheating Temperature setting range

+70℃~+200℃

Heating time: about 80 minutes from room temperature to +200 ℃

Low temperature liquid tank liquid precooling

Temperature setting range

﹣80℃~0℃

Cooling time: about 90 minutes from room temperature to -80℃

Air conditioning room environment Temperature < 25℃; Relative humidity < 65%RH
Temperature recovery time

<5min

Temperature conversion time

<10sec

Standard

GB2423.1
GB2423.2
GB2423.22

GB10589-89
GB11158-89
GB10592-89

GJB150.3-86
GJB150.4-86

IEC68-2-14

MIL-STD-883

MIL-STD-810F

MIL-STD-202F